Home » WEF Discussion Forums » Laboratory Management & Technical Issues » Zeroing spectophotometer for COD analysis


Zeroing spectophotometer for COD analysisExpand / Collapse
Author
Message
Posted 5/2/2008 5:50:19 PM
Forum Member

Forum MemberForum MemberForum MemberForum MemberForum MemberForum MemberForum MemberForum Member

Group: Forum Members
Last Login: 6/16/2008 4:32:34 PM
Posts: 38, Visits: 259
Our lab is developing a colorimetric procedure for detemination of chemical oxygen demand, after many years of titration. We are using Hach tubes containing digestion reagents, and a Hach DR 5000 spectrophotometer. My question regards the proper means of zeroing the instrument absorbance prior to calibration.

The information from Hach suggests the instrument should be zeroed with a digested blank (which I would consider a method blank). Standard Methods 5220 D suggests zeroing with deionized water. For some other procedures (e.g., total phosphate) we have been zeroing with a mixture of digestion solution and deionized water (some would call it a reagent blank, although others hold that a reagent blank can contain no matrix - i.e., no DI water).

Any feedback, opinions, or information regarding what your labs are doing would be helpful. Thanks.

Post #9050
Posted 5/5/2008 8:21:21 AM


Supreme Being

Supreme BeingSupreme BeingSupreme BeingSupreme BeingSupreme BeingSupreme BeingSupreme BeingSupreme Being

Group: Forum Members
Last Login: 7/1/2008 2:01:12 PM
Posts: 187, Visits: 1,757
First off, you have to notice that SM gives you two methods for analyzing COD.  The first and most common is what you are using now that goes from about 50-100 up to about 1000mg/L COD.  Use an undigested blank as your zero initially.  SM says that once you determine your MB to be consistently clean you can use a digested blank as your zero. 

The part you see that says to use reagent water as your blank is for the low level reagents.  You measure this one at 420nm, as opposed to 600nm, and the curve is backwards from what you expect.  The color will actually get lighter the higher the concentration.  This one has a range of 10-150 mg/L of COD.

For everything, except low level COD's, we zero with a true method blank carried through the entire analytical procedure.

David

Post #9055
Posted 5/6/2008 3:22:23 PM
Forum Member

Forum MemberForum MemberForum MemberForum MemberForum MemberForum MemberForum MemberForum Member

Group: Forum Members
Last Login: 6/12/2008 9:19:35 AM
Posts: 37, Visits: 188
I have always zeroed the spec with the digested blank, for both methods - low and high range.  Never had a problem with the calibration curves doing it this way. 

Brandi
Post #9071
« Prev Topic | Next Topic »


Reading This TopicExpand / Collapse
Active Users: 1 (1 guest, 0 members, 0 anonymous members)
No members currently viewing this topic.

PermissionsExpand / Collapse

All times are GMT -5:00, Time now is 12:37am

Powered By InstantForum.NET v4.0.0 © 2008
Execution: 0.109. 10 queries. Compression Enabled.